Tóm tắt tài liệu kỹ thuật
Testomat 2000 FE
Engineering summary from PDF text extraction for Testomat 2000 FE. Verify every value with the OEM datasheet.
Summary
Product identification
The Testomat 2000® Fe is a process photometer designed for monitoring dissolved iron content (II)+(III).
Intended use
This instrument is a wet-chemical online monitor for the colorimetric analysis of dissolved iron. It is suitable for applications such as monitoring de-ironing plants or process and drinking-water flows. The intended use requires adherence to the specified performance limits and medium requirements.
Operating limits
- Pressure range: 0.3 to 8 bar (factory setting). Can be adjusted for 0.3 to 1 bar by removing the flow controller valve body.
- Water temperature: 10°C to 40°C. A KCN type cooler is recommended for temperatures above 40°C.
- pH range: 4 – 10.5.
- Influencing ions: Copper above 0.01 mg/l and aluminum above 0.1 mg/l can affect the color reaction.
Feed water
- Water inlet pressure: 0.3 bar to 8 bar. Avoid strong pressure fluctuations.
- Inlet hose: Opaque plastic hose 6/4 x 1 (external diameter 6 mm, internal diameter 4 mm, wall thickness 1 mm), maximum length 5 m.
- Pump capacity (for 0.3-1 bar operation): 25 to 35 litres/hour, resistant to the measured medium.
Product water
- Measurement range: 0 – 1.0 mg/l dissolved iron (II)+(III).
- Range I: 0 - 0.65 mg/l, resolution: 0.01 mg/l
- Range II: 0.7 – 1.0 mg/l, resolution 0.1 mg/l
- Analysis time: Approximately 7 minutes (absolute measuring time without flushing).
Electrical
- Mains voltage: Connectable to 24 V / 115 V / 230 V AC. Refer to the rating plate for the specific voltage.
- Relay outputs: Maximum load capacity of 4 A at 240 VAC.
- Input/Output connections: Designed for volt-free contacts. Do not connect external voltage.
Mechanical / hydraulic
- Water inlet connection: Standard plug connector for opaque plastic hoses 6/4 x 1.
- Water outlet connection: For outlet hose with an internal diameter of 12 mm.
- Reagent storage: 500 ml reagent bottles. Uses reagents FE2005A and FE2005B.