Back to Veolia Filtration
Veolia logo

Datasheet summary

Memtrex MP-S

Engineering summary from PDF text extraction for Memtrex MP-S. Verify every value with the OEM datasheet.

Summary

Product identification

The Memtrex MP-S are pleated filters designed for final sterile filtration of biological and pharmaceutical products. They feature a double-layer polyethersulfone (PES) membrane providing low protein-binding, high throughput, and broad chemical compatibility. These filters are available with an absolute pore size micron rating of 0.2 µm.

Typical Applications

Memtrex MP-S filters are specifically designed for sterile filtration of pharmaceutical and biological products. Typical applications include:

  • Final Filtration of Drugs and Biologicals
  • Final Filtration of Water for Injection
  • Final Filtration of Pure Water for Dialysis
  • LVPs, SVPs
  • Diagnostics
  • Tissue Culture Media
  • Vaccines

Materials of Construction

  • Media: 2 layers of Hydrophilic Polyethersulfone
  • Support Layers: Polypropylene Microfiber
  • Core and Cage: Polypropylene
  • End caps and Adapters: Polypropylene with Stainless Steel Insert

Dimensions

  • Nominal O.D.: 2.75” (70 mm)
  • Nominal I.D.: 1.25” (31 mm)
  • Effective Filtration Area: 6.5 ft² (0.60 m²)

Operating limits

  • Max. Forward Differential Pressure: 60 psi (4.1 bar)
  • Max. Reverse Differential Pressure: 30 psi (2.1 bar)
  • Max. Operating Temperature: 180ºF (82ºC) at 10 psid (0.7 bar) in water

Integrity Testing

  • Air diffusion per 10" module after saturation with clean water <12 cc/min at 40 psig (2.76 bar).

Flow Performance

The flow performance in clean water is listed in the original document, based on a 10" length filter. See the official PDF for exact tabulated values.

Additional Information

  • Memtrex MP-S filters may be autoclaved or in situ steam sterilized (up to 257°F [125°C], 30-minute cycles) for a maximum accumulated exposure of 10 hours. Compatible chemical agents can also be used for sanitization.
  • The materials meet U.S. FDA requirements for food contact under 21 CFR and meet the test criteria for USP class VI-121°C Plastics.
  • Aqueous extracts from Memtrex MP-S filters contain less than 0.25 EU/ml, and the filters typically exhibit low levels of non-volatile residues.
  • Users should carefully verify chemical compatibility for specific applications.

Ordering Information

The ordering information for Memtrex MP-S includes specifications for:

  • Type: MMP-S
  • Absolute Micron Rating: 92 = 0.2 μm
  • Nominal Cartridge Length: 1 = 10 inch (25 cm), 2 = 20 inch (51 cm), 3 = 30 inch (76 cm), 4 = 40 inch (102 cm)
  • End #1 Adapter: Q = 222 O-Ring Stainless Steel Support Ring, Z = 226 O-Ring Stainless Steel Support Ring
  • End #2 Adapter: A = Open End Gasket, B = 120 O-Ring, C = 213 O-Ring, G = Closed End Cap, H = Fin Adapter
  • Elastomer Material: S = Silicone
  • Grade: S = Sterilizing

See the official PDF for exact tabulated values and a complete list of ordering options.


Disclaimer: This summary is based on the extracted text and may not include all figures, footnotes, or the latest revisions from the original OEM PDF datasheet. For contractual data, please refer to the specific OEM PDF revision used on your project.