
ملخص ورقة البيانات
Titromat M1
Engineering summary from PDF text extraction for Titromat M1. Verify every value with the OEM datasheet.
Summary
Product identification
Titromat® M1 is used for automatically determining and monitoring carbonate hardness in low hardness measuring ranges.
Applications
- Corrosion monitoring in boiler feed water.
- Residual alkalinity after decarbonization (e.g., breweries).
Operating limits
Measuring range
- 0.05 - 1.00 °dH (resolution 0.025 °dH)
- 0.09 - 1.80 °f (resolution 0.045 °f)
Feed water
Operating pressure
- 1 to 8 bar (15 to 116 psi)
- 0.3 to 1 bar (5 to 15 psi) (after removing the valve body from the regulator/filter housing)
Water inlet
- Opaque hose with 6 mm external diameter / 4 mm internal diameter.
Water outlet
- Hose with 6 mm external diameter / 4 mm internal diameter.
Water temperature
- 10 to 40 °C (50 to 104°F).
Electrical
- Power supply: 230–240 VAC, 115 VAC, or 24 VAC, all 50–60 Hz.
- Power consumption: Max. 30 VA.
- Protection class: I.
Mechanical / hydraulic
- Degree of protection: IP 65.
- Ambient temperature: 10–45 °C (50 - 104°F).
- Dimensions (W x H x D): 380 x 480 x 280 mm.
- Weight: Approx. 9.5 kg.
- Menu languages: German, English, French (others upon request).
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.
Official datasheet (PDF)
أُعد المحتوى من مراجع تقنية عامة للبحث والمقارنة الأولية. لا يغني الملخص عن ورقة بيانات حالية معتمدة من الشركة المصنعة. تحقق من الإصدارات وحدود التصميم قبل الاستخدام.