العودة إلى Testomat
Testomat logo

ملخص ورقة البيانات

Testomat 2000 Polymer

Engineering summary from PDF text extraction for Testomat 2000 Polymer. Verify every value with the OEM datasheet.

Summary

Product identification

The Testomat 2000® Polymer is a process photometer and wet-chemical online monitor for polyacrylate content.

Application

  • Monitoring of conditioning agents in cooling and heating circuits.

Measuring range

  • 0 to 50 mg/l (ppm) with a factor set to 1.00.
  • Customer-specific ranges, e.g., 0.0-50.0 mg/l.

Customization is necessary due to the range of polyacrylates measurable.

Technical Data

  • Power supply: 230–240 VAC, 115 VAC, 24 VAC (50–60 Hz)
  • Power consumption: max. 30 VA
  • Protection class: I
  • Degree of protection: IP 65
  • Ambient temperature: 10–45 °C (50 - 104°F)
  • Dimensions (W x H x D): 380 x 480 x 280 mm
  • Weight: approx. 9.5 kg
  • Menu languages: German, English (others upon request)

Operating limits

Mains water supply

  • Operating pressure: 1 to 8 bar (15 to 116 psi) or 0.3 to 1 bar (5 to 15 psi) after removing the valve body from the regulator/filter housing.
  • Water temperature: 10 to 40 °C (50 to 104°F)

Connections

  • Water inlet: Opaque hose with 6 mm external diameter / 4 mm internal diameter.
  • Water outlet: Hose with 6 mm external diameter / 4 mm internal diameter.

Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.