العودة إلى Testomat
Testomat logo

ملخص ورقة البيانات

Testomat 2000 CLO2

Engineering summary from PDF text extraction for Testomat 2000 CLO2. Verify every value with the OEM datasheet.

Summary

Product identification

The Testomat 2000® ClO2 is a wet-chemical online instrument for monitoring chlorine dioxide (ClO₂) content.

Application

  • Disinfectant monitoring for drinking water and process water.

Operating limits

  • Ambient temperature: 10–45 °C (50 - 104°F)
  • Water temperature: 10 to 40 °C (50 to 104°F)

Feed water

  • Operating pressure:
    • 1 to 8 bar (15 to 116 psi)
    • 0.3 to 1 bar (5 to 15 psi), after removing the valve body from the regulator/filter housing.
  • Water inlet: Opaque hose with 6 mm external diameter / 4 mm internal diameter.

Product water

  • Measurement range: 0–4.7 mg/l (ppm)
  • Resolution:
    • 0.02 mg/l (ppm) in the range 0 - 1.88 mg/l (ppm)
    • 0.2 mg/l (ppm) in the range 1.9 - 4.7 mg/l (ppm)
  • Water outlet: Hose with 6 mm external diameter / 4 mm internal diameter.

Electrical

  • Power supply: 230–240 VAC, 115 VAC, 24 VAC (all 50–60 Hz)
  • Power consumption: max. 30 VA
  • Protection class: I
  • Degree of protection: IP 65

Mechanical / hydraulic

  • Dimensions (W x H x D): 380 x 480 x 280 mm
  • Weight: approx. 9.5 kg
  • Menu languages: German, English, French (others upon request)
  • Option: Booster pump (see accessories)