Product identification

  • Product Name: MicroPES® Type TF10 iv
  • Intended Use: Infusion therapy
  • Membrane Type: Capillary membrane
  • Document Revision: Sheet No. 425/9061/000 of 03/16, supersedes 425/9061/000 of 11/13

Chemical composition

  • Polymer: Polyethersulfone

Mechanical / hydraulic

  • Wall thickness: 100 ± 25 µm
  • Inner diameter: 300 ± 40 µm
  • Tensile strength: ≥ 50 cN
  • Elongation at break: ≥ 30 %

Membrane performance characteristics

  • Transmembrane flow (water, 25°C): ≥ 35 ml/[min x cm² x bar]
  • Maximum pore size (calculated from BP): 0.5 ± 0.1 µm
  • Bubble point (Isopropylalcohol, 23°C): 1.22 +0.31 / -0.20 bar

Available configurations

  • Single-thread spools

Disclaimer: This summary is based on extracted text and may not include all figures, footnotes, or the latest revisions from the official PDF datasheet. Contractual data must always match the specific OEM PDF revision used on the project.