Summary
Product identification
- Product Type: Capillary Ultrafiltration Module
- Model: HYDRAcap®60
- Membrane Polymer: Hydrophilic Polyethersulfone
- Nominal Membrane Area: 500 ft² (46 m²)
- Fiber Dimensions:
- ID 0.031” (0.8 mm)
- OD 0.055” (1.4 mm)
- Pore size: 0.02 micron
- Certifications: NSF61, CDPH, and ETV-NSF Verification
Performance
- Filtrate Flow: 12.1 – 29.5 gpm (2.7 – 6.7 m³/h)
- Filtrate Turbidity: ≤ 0.07 NTU
- Virus removal: ≥ 4 log
- Bacteria removal: ≥ 4 log
Operating limits
- Typical Filtrate Flux Range: 35 – 85 gfd (59 – 145 l/m²/h)
- Maximum Applied Feed Pressure: 73 psig (5 bar) (at 68°F (20°C))
- Maximum Transmembrane Pressure: 20 psig (1.4 bar)
- Maximum Backwash Transmembrane Pressure: 20 psig (1.4 bar)
- Instantaneous Chlorine Tolerance: 100 ppm (for 15 minutes or less)
- Instantaneous Hydrogen Peroxide Tolerance: 200 ppm (for 15 minutes or less)
- Maximum Chlorine Exposure: 200,000 ppm-hrs
- Maximum Instantaneous Feed Turbidity: 100 NTU
- Maximum Operating Temperature: 104 °F (40 °C)
- pH Operating Range: 4.0 - 10.0
- Cleaning pH Range: 1.5 – 13.0
- Operating Mode: Inside to Outside Filtration (Direct flow or Crossflow)
Process Conditions
- Backwash Flux: 100 – 150 gfd (170 – 255 l/m²/h)
- Backwash Duration: 30 – 60 seconds
- Backwash Frequency: 20 – 60 minutes
- Chemical Enhanced Backwash Frequency: 0 – 4 times per day
- Chemical Enhanced Backwash Duration: 1 – 30 minutes
- Disinfection Chemicals: NaOCl, H₂O₂, ClO₂ or NH₂Cl
- Cleaning Chemicals: NaOH, HCl, H₂SO₄, or Citric Acid
Mechanical / hydraulic
- Dimensions:
- A: 63 inches (1600 mm)
- B: 66 1/8 inches (1680 mm)
- C: 67 1/4 inches (1708 mm)
- Pipe connections: 2” Victaulic
- Dry Weight: 80 lbs (36.3 kg)
- Wet Weight: 155 lbs (70.3 kg)
- Note: Weight stated is shipping weight including 1L of a 0.95% solution of sodium bisulfite preservative.