
Tóm tắt tài liệu kỹ thuật
Titromat M2
Engineering summary from PDF text extraction for Titromat M2. Verify every value with the OEM datasheet.
Summary
Product identification
Titromat® M2
This instrument is used for automatically determining and monitoring the carbonate hardness for low hardness measuring ranges.
Application
- Corrosion monitoring in boiler feed water
- Residual alkalinity after decarbonization (e.g., breweries)
Operating limits
Measuring range
- 0.05 - 2.00 °dH (resolution 0.05 °dH)
- 0.09 - 3.60 °f (resolution 0.09 °f)
Environmental conditions
- Ambient temperature: 10–45 °C
- Water temperature: 10 to 40 °C
Feed water
Hydraulic connection
- Operating pressure:
- 1 to 8 bar (15 to 116 psi)
- 0.3 to 1 bar (5 to 15 psi) (after removing the valve body from the regulator/filter housing)
- Inlet hose: Opaque hose with 6 mm outer diameter / 4 mm inner diameter
- Outlet hose: Hose with 6 mm outer diameter / 4 mm inner diameter
Electrical
- Supply voltage: 230–240 VAC, 115 VAC, 24 VAC (all 50–60Hz)
- Power consumption: max. 30 VA
- Protection class: I
- Type of protection: IP 65
Mechanical / hydraulic
- Dimensions (W x H x D): 380 x 480 x 280 mm
- Weight: approx. 9.5 kg
- Menu languages: German, English, French (others on request)
Disclaimer: The extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.
Official datasheet (PDF)
Nội dung tổng hợp từ tài liệu kỹ thuật công khai nhằm tra cứu và so sánh sơ bộ. Bản tóm tắt không thay thế bảng dữ liệu hiện hành được nhà sản xuất chứng nhận. Kiểm tra phiên bản và giới hạn thiết kế trước khi sử dụng.