Quay lại Testomat
Testomat logo

Tóm tắt tài liệu kỹ thuật

Titromat M2

Engineering summary from PDF text extraction for Titromat M2. Verify every value with the OEM datasheet.

Summary

Product identification

Titromat® M2

This instrument is used for automatically determining and monitoring the carbonate hardness for low hardness measuring ranges.

Application

  • Corrosion monitoring in boiler feed water
  • Residual alkalinity after decarbonization (e.g., breweries)

Operating limits

Measuring range

  • 0.05 - 2.00 °dH (resolution 0.05 °dH)
  • 0.09 - 3.60 °f (resolution 0.09 °f)

Environmental conditions

  • Ambient temperature: 10–45 °C
  • Water temperature: 10 to 40 °C

Feed water

Hydraulic connection

  • Operating pressure:
    • 1 to 8 bar (15 to 116 psi)
    • 0.3 to 1 bar (5 to 15 psi) (after removing the valve body from the regulator/filter housing)
  • Inlet hose: Opaque hose with 6 mm outer diameter / 4 mm inner diameter
  • Outlet hose: Hose with 6 mm outer diameter / 4 mm inner diameter

Electrical

  • Supply voltage: 230–240 VAC, 115 VAC, 24 VAC (all 50–60Hz)
  • Power consumption: max. 30 VA
  • Protection class: I
  • Type of protection: IP 65

Mechanical / hydraulic

  • Dimensions (W x H x D): 380 x 480 x 280 mm
  • Weight: approx. 9.5 kg
  • Menu languages: German, English, French (others on request)

Disclaimer: The extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.