Summary
Product identification
- Product: Testomat 808 SiO2
- Application: Sterilisation of hospitals, silica monitoring
Operating limits
- Silica (SiO₂) measuring range: 0.3 - 1.2 ppm
- Ambient temperature: 10 – 40 °C (50 - 104°F)
- Water temperature: 10 to 40 °C (50 to 104°F)
Feed water
- Operating pressure:
- 0.3 – 1 bar (0.3 x 10⁵ to 1 x 10⁵ Pa)
- 1 - 4 bar (1 x 10⁵ to 4 x 10⁵ Pa)
- (Note: A pressure reducer is recommended from 4 to 8 bar)
- Water inlet hose: 6 mm external diameter / 4 mm internal diameter
- Water outlet hose: 6 mm external diameter / 4 mm internal diameter
Electrical
- Power supply: 24V / 115V / 230V, 50 – 60 Hz (depending on configuration)
- Instrument protection: T0,1 A (230-240V); T0,2 A (115V); T0,8 A (24V)
- Mains protection for consumers: max. 4 A (N, L)
- Power consumption: max. 16 VA (without external load)
- Protection class: I
- Current interface: Output of defined values (5, 8, 11, 14, 17, 20 mA); max. load 500 Ohms
- Contact load relay: 230V / 4A AC ohm resistive load
Mechanical / hydraulic
- Dimensions (W x H x D): 364 x 314 x 138 mm
- Dimensions with side pocket: 442 x 314 x 138 mm
- Weight: 4350 g
- Degree of protection: IP 54
Conformity
- EN 285:2006
- EN 61000-6-2
- EN 61000-6-4
- EN 61010-1
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.