
Tóm tắt tài liệu kỹ thuật
Testomat 2000 Polymer
Engineering summary from PDF text extraction for Testomat 2000 Polymer. Verify every value with the OEM datasheet.
Summary
Product identification
The Testomat 2000® Polymer is a process photometer and wet-chemical online monitor for polyacrylate content.
Application
- Monitoring of conditioning agents in cooling and heating circuits.
Measuring range
- 0 to 50 mg/l (ppm) with a factor set to 1.00.
- Customer-specific ranges, e.g., 0.0-50.0 mg/l.
Customization is necessary due to the range of polyacrylates measurable.
Technical Data
- Power supply: 230–240 VAC, 115 VAC, 24 VAC (50–60 Hz)
- Power consumption: max. 30 VA
- Protection class: I
- Degree of protection: IP 65
- Ambient temperature: 10–45 °C (50 - 104°F)
- Dimensions (W x H x D): 380 x 480 x 280 mm
- Weight: approx. 9.5 kg
- Menu languages: German, English (others upon request)
Operating limits
Mains water supply
- Operating pressure: 1 to 8 bar (15 to 116 psi) or 0.3 to 1 bar (5 to 15 psi) after removing the valve body from the regulator/filter housing.
- Water temperature: 10 to 40 °C (50 to 104°F)
Connections
- Water inlet: Opaque hose with 6 mm external diameter / 4 mm internal diameter.
- Water outlet: Hose with 6 mm external diameter / 4 mm internal diameter.
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.
Official datasheet (PDF)
Nội dung tổng hợp từ tài liệu kỹ thuật công khai nhằm tra cứu và so sánh sơ bộ. Bản tóm tắt không thay thế bảng dữ liệu hiện hành được nhà sản xuất chứng nhận. Kiểm tra phiên bản và giới hạn thiết kế trước khi sử dụng.