Tóm tắt tài liệu kỹ thuật
Testomat 2000 CLT self clean
Engineering summary from PDF text extraction for Testomat 2000 CLT self clean. Verify every value with the OEM datasheet.
Summary
Product identification
Testomat 2000® CLT self clean - a process photometer with an additional pump for dosing detergent.
Application
- Disinfectant monitoring for drinking water and process water
- Medical technology (dialysis)
Operating limits
- Ambient temperature: 10–45 °C (50 - 104°F)
- Water temperature: 10 to 40 °C (50 to 104°F)
Feed water
- Operating pressure: 1 to 8 bar (15 to 116 psi) OR 0.3 to 1 bar (5 to 15 psi) after removing the valve body from the regulator/filter housing.
- Water inlet: Opaque hose with 6 mm external diameter / 4 mm internal diameter.
Product water
- Measuring range (Total chlorine): 0.00 - 0.99 mg/l and 1.0 - 2.5 mg/l
- Water outlet: Hose with 6 mm external diameter / 4 mm internal diameter.
Electrical
- Power supply: 230–240 VAC, 115 VAC, 24 VAC (all 50–60Hz)
- Power consumption: max. 30 VA
- Protection class: I
- Degree of protection: IP 65
Mechanical / hydraulic
- Dimensions (W x H x D): 380 x 480 x 280 mm
- Weight: approx. 9.5 kg
- Menu languages: German, English, French (others upon request)
- Booster pump: Option available (see accessories)
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.