Quay lại The AQL Ozone family
Aeroqual logo

Tóm tắt tài liệu kỹ thuật

S900

Engineering summary from PDF text extraction for S900. Verify every value with the OEM datasheet.

Summary

Product identification

  • Series: 900 Transmitter / Controller
  • Sensor Heads: Ozone (Low, High, Leak), VOC, Ammonia, CO

Operating limits

  • Operating temperature range (Sensor head and base unit): -5°C to 50°C; 23°F to 122°F
  • Relative humidity limit (Sensor head and base unit): 95% maximum

Ozone Sensor Heads

  • Low Measurement range: 0.000 - 0.500 ppm
  • Low Accuracy: < 0.008 ppm (0 - 0.100 ppm); ± 10% (0.100 – 0.500 ppm)
  • Low Resolution: 0.001 ppm
  • Low T90 Response: < 60 seconds
  • High Measurement range: 0.50 - 20.00 ppm
  • High Accuracy: ± 10% (0.20 - 2.00 ppm); ± 15% (2.00 - 20.00 ppm)
  • High Resolution: 0.01 ppm
  • High T90 Response: < 35 seconds
  • Leak Measurement range: 0.00 - 50.00 ppm
  • Leak Accuracy: ± 20%, 15 seconds after Reset
  • Leak Resolution: 0.01 ppm
  • Leak T90 Response: < 10 seconds
  • Sensor type: Gas-sensitive semiconductor

VOC Sensor Head Specification

  • Measurement range: 0 to 400 ppm toluene
  • Accuracy: < +/- 10 ppm (0 – 200 ppm); < +/- 10% ( > 200 ppm)
  • T90 response: < 60 s
  • Sensor type: Gas-sensitive semiconductor
  • Operating temperature range: -20°C to 60°C
  • Operating relative humidity range: 5 to 95% non-condensating

Ammonia Sensor Head Specification

  • Measurement range: 0 to 1000 ppm
  • Accuracy: < +/- 5 ppm (0 – 100 ppm); < +/- 10% (100 – 1000 ppm)
  • T90 response: < 60 s
  • Sensor type: Gas-sensitive semiconductor
  • Operating temperature range: -20°C to 60°C
  • Operating relative humidity range: 5 to 95% non-condensating

CO Sensor Head Specification

  • Measurement range: 0 to 2000 ppm carbon monoxide
  • Accuracy: < +/- 10 ppm in the range 0 to 400 ppm
  • T90 response: < 150s
  • Sensor type: Gas-sensitive semiconductor
  • Operating temperature range: 0°C to 70°C
  • Operating relative humidity range: 5 to 95% non-condensating

Electrical

  • Power (user supplied): Regulated 12 VDC, 800 mA
  • Analogue output: 4-20 mA (opto-isolated), 10-30 V
  • External signal type: Transistor Output
  • External signals: 4
  • External signal functions: Low Alarm, High Alarm, Control & Diagnostics
  • Inputs: Standby toggle
  • Communication: RS485
  • RS485 protocol: Aeroqual proprietary product

Mechanical / hydraulic

  • Removable / replaceable sensor head: Yes
  • Enclosure rating: IP20 & NEMA 1 equivalent
  • Enclosure size (with base unit): 64 H x 130 Ø (mm); 2½ H x 5⅛Ø (in)
  • Mounting: Screw fixture
  • Weight: < 200 g; < 7 oz

Controls and Configuration

  • Jumpers: J1, J2, J3 termination resistors
  • Connectors: Screw
  • ID: 1 (Default); User configurable from 1 - 255
  • Alarm set points (2): User configurable
  • Control set point: User configurable

Approvals

  • Base Unit: Part 15 of the FCC Rules, EN 50082-1: 1997, EN 50081-1: 1992
  • CO Sensor Head: UL 2034, BS 7860

Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.

Official datasheet (PDF)

Lenntech datasheet mirror

Nội dung tổng hợp từ tài liệu kỹ thuật công khai nhằm tra cứu và so sánh sơ bộ. Bản tóm tắt không thay thế bảng dữ liệu hiện hành được nhà sản xuất chứng nhận. Kiểm tra phiên bản và giới hạn thiết kế trước khi sử dụng.