Tóm tắt tài liệu kỹ thuật
PuriSys CPF201
Engineering summary from PDF text extraction for PuriSys CPF201. Verify every value with the OEM datasheet.
Summary
Product identification
PuriSys CPF 201 - Compact pH Electrode
Application:
- Boiler feed water
- Power plant applications
- Pharmaceutical water for injection (WFI)
Operating limits
- Process temperature: max. 75 °C / 167 °F
- Process pressure: max. 3.45 bar / 50 psi
- Medium conductivity: < 100 µS/cm
- Drift: < 0.02 pH/week at 0.5 µS/cm at 25 °C / 77 °F
Feed water
- Medium conductivity: < 100 µS/cm
Product water
- Designed for use in pure and ultrapure water applications.
Operating parameters
- pH measurement range: 0 ... 14 pH
- Temperature measurement range: 2 ... 75 °C / 36 ... 167 °F
- Response time (t₉₀): < 10 s
- Flow rate: 50 ... 250 ml/min (Optimum: 150 ml/min)
Electrical
- Temperature compensation sensor: Pt 100 or Pt 1000 RTD
Mechanical / hydraulic
- Electrode shaft material: Stainless steel SS 316, Viton®
- Reference junction material: Teflon®
- Glass bulb: pH membrane glass
- Reference system: KCl/AgCl
- Flow assembly material: Stainless steel SS 316, Viton®
- Inlet / outlet ports: Stainless steel SS 316
- Process connection:
- OD 6 mm stainless steel compression fitting
- ¼" stainless steel compression fitting
- Cable connection: TOP 68 quick disconnect, water tight connector
- Ingress protection: IP 68 / NEMA 6 (TOP 68 plug-in system)
- Weight:
- Electrode: approx. 213 g
- Complete on-line unit incl. mounting brackets: approx. 2275 g
Scope of delivery
- 1 CPF 201 electrode
- 1 flow assembly
- 1 PVC mounting plate incl. mounting brackets
- Technical Information TI 366C/07/en
Accessories
- Measuring cable CPK 9 (version without PM)
- Buffer solutions (pH 4.0, pH 7.0)
- O-ring set
- Flow assembly lid
- Compression fittings (¼", OD 6 mm)
- Mounting plate
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.
Official datasheet (PDF)
Nội dung tổng hợp từ tài liệu kỹ thuật công khai nhằm tra cứu và so sánh sơ bộ. Bản tóm tắt không thay thế bảng dữ liệu hiện hành được nhà sản xuất chứng nhận. Kiểm tra phiên bản và giới hạn thiết kế trước khi sử dụng.