Summary

Product identification

  • Product name: Memosens CPS42E
  • Product type: ORP sensor for process technology
  • Technology: Digital with Memosens 2.0 technology

Application

  • Media: Low conductivity, or high proportion of organic solvents or alcohol.
  • Industries: Chemical industry, organic chemicals, power stations, laboratory measurements.
  • Hazardous Areas: ATEX, IECEx, CSA C/US, NEPSI, Japan Ex and INMETRO approvals for Zone 0, Zone 1, and Zone 2.

Operating limits

  • Pressure: Up to 11 bar (159.5 psi) absolute when counterpressure is applied.
  • Temperature (Process): -15 to 135 °C (5 to 275 °F).
  • Temperature (Ambient): Not specified below 0 °C (32 °F) due to risk of frost damage. Storage temperature: 0 to 50 °C (32 to 122 °F).

Feed water

  • Conductivity: Minimum 5 μS/cm (Reference system SB, with minimized flow; pressure and temperature must remain constant).

Product water

  • ORP Measurement Range: -1500 to 1500 mV.

Electrical

  • Power Supply: Via Memosens data cable (CYK10 or CYK20) connected to a Memosens transmitter.
  • Data Transmission: Digital, non-contact, inductive signal transmission.

Mechanical / hydraulic

  • Reference System: Ag/AgCl reference lead with 3M KCl liquid electrolyte.
  • Junction: Ceramic junction with specified KCl flow rate.
  • Temperature Sensor: Integrated NTC 30K.
  • Material:
    • Sensor shaft: Glass
    • ORP measuring element: Platinum
    • O-ring: FKM
    • Process coupling: PPS fiber-glass reinforced
    • Plug-in head: Memosens plug-in head, pressure resistance 16 bar (232 psi) relative.
  • Dimensions: Various installed lengths available (120 mm, 225 mm, 360 mm, 425 mm).
  • Weight: Ranges from 40 g to 100 g depending on installed length.
  • Process Connection: Pg 13.5.
  • Degree of Protection: IP 68.
  • Installation Angle: Minimum 15° from horizontal is required to ensure electrolyte contact.

Certificates and approvals

  • CE Mark: Complies with EU directives.
  • Ex Approvals: ATEX, IECEx, NEPSI, CSA C/US, Japan Ex, INMETRO.
  • Other: EAC certification.