Tóm tắt tài liệu kỹ thuật
Micropilot FMR10B
Engineering summary from PDF text extraction for Micropilot FMR10B. Verify every value with the OEM datasheet.
Summary
Product identification
- Product: Micropilot FMR10B
- Type: Free-space radar level sensor
- Document: TI01805F/00/EN/02.24-00, dated 2024-12-09
Application
- Continuous, non-contact level measurement of liquids and bulk solids.
- Flow measurement in open channels or weirs with totalizer.
Operating limits
- Process Temperature: –40 to 60 °C (–40 to 140 °F)
- Process Pressure: –1 to 3 bar gauge (–14 to 43 psi gauge)
- Operating Height: Up to 5,000 m (16,404 ft) above sea level.
- Ambient Temperature Range: –40 to +60 °C (–40 to +140 °F)
- Storage Temperature: –40 to +80 °C (–40 to +176 °F)
Feed water / Medium
- Relative Permittivity (Liquids): εr ≥ 1.8 (contact Endress+Hauser for lower values)
- Relative Permittivity (Bulk Solids): εr ≥ 1.6
Product water / Measurement
- Maximum Measuring Range: 10 m (33 ft)
- Accuracy:
- Liquids: Up to ± 5 mm (0.2 in) for distances > 0.25 m (0.82 ft).
- Solids: Up to ± 10 mm (0.39 in) for distances > 0.8 m (2.6 ft).
- Resolution:
- Current output: < 1 µA
- Digital: 1 mm (0.04 in)
- Non-repeatability: ≤ 1 mm (0.04 in)
- Warm-up time: ≤ 30 s
Electrical
- Supply Voltage: 12 to 30 VDC
- Power Consumption: Maximum current limited to 500 mA.
- Output Signal: 4 to 20 mA
- Operating modes: 4 to 20.5 mA, NAMUR NE 43 (3.8 to 20.5 mA factory setting), US mode (3.9 to 20.5 mA).
- Signal on alarm (current output): Max. alarm 21.5 to 23 mA, Min. alarm < 3.6 mA (factory setting).
- Overvoltage protection: 1,000 V wire to ground for DC power supply and IO lines.
- Overvoltage Category: II
Mechanical / Hydraulic
- Degree of Protection: IP66/68, NEMA Type 4X/6P (24 h at 1.83 m (6.00 ft) under water).
- Antenna: 40 mm (1.5 in)
- Weight (Device with 10 m cable): Approx. 0.7 kg (1.5 lb)
- Materials:
- Antenna end process connection: PVDF
- Seals: EPDM
- Housing: PBT/PC
- Connecting cable: PVC
Communication
- Wireless Technology: Bluetooth®
- Supported Operating Tools:
- Free iOS/Android app SmartBlue
- DeviceCare software (version 1.07.07 and higher)
- FieldXpert SMT70/SMT77
Environmental
- Vibration Resistance: Guaranteed for 5 to 2,000 Hz: 1.25 (m/s²)²/Hz, ~5 g.
- Electromagnetic Compatibility (EMC): As per EN 61326 series and NAMUR recommendation EMC (NE21). Maximum measured error during EMC testing: < 0.5 % of the span.
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.
Official datasheet (PDF)
Nội dung tổng hợp từ tài liệu kỹ thuật công khai nhằm tra cứu và so sánh sơ bộ. Bản tóm tắt không thay thế bảng dữ liệu hiện hành được nhà sản xuất chứng nhận. Kiểm tra phiên bản và giới hạn thiết kế trước khi sử dụng.