العودة إلى 3M Liqui-Flux Ultrafiltration
3M logo

ملخص ورقة البيانات

DIAPES HF800

PDF datasheet mirror for DIAPES HF800 (3M Liqui-Flux Ultrafiltration). Verify with OEM before design.

Summary

Product identification

  • Product Name: DIAPES® Capillary membrane
  • Type: HF800*
  • Document Revision: 079/0107/000 of 03/16 (supersedes 079/0107/000 of 11/13)

Chemical composition

  • Polymer: Polyethersulfone

Mechanical / hydraulic

  • Wall thickness: 30 µm ± 5 µm
  • Inner diameter: 200 µm ± 15 µm
  • Available make-up configurations: Bundles (customer specified), with P.E.T.®

Membrane performance characteristics

  • Hydraulic permeability: approx. 38 ml/[h x m² x mmHg] (ultrafiltration rate for Albumin solution) (QF = 30 ml/min x m²)
  • Sieving coefficients (QF = 30 ml/min x m²):
    • Cytochrome c: approx. 0.63
    • Albumin: approx. 0.01

Disclaimer: This summary is based on extracted text and may not include all figures, footnotes, or the latest revisions from the original PDF. Contractual data must always match the specific OEM PDF revision used on the project.