Summary

Product identification

  • Product name: Testomat 808 SiO₂
  • Application: Sterilization of hospitals, silica monitoring
  • Compliance: EN 285:2006

Operating limits

  • Ambient temperature: 10 – 40 °C (50 - 104°F)
  • Water temperature: 10 to 40 °C (50 to 104°F)

Feed water

  • Operating pressure:
    • 0.3 – 1 bar (0.3 x 10⁵ to 1 x 10⁵ Pa)
    • 1 - 4 bar (1 x 10⁵ to 4 x 10⁵ Pa)
    • A pressure reducer (special accessory) should be used from 4 to 8 bar range.
  • Water inlet: Opaque hose with 6 mm external diameter / 4 mm internal diameter
  • Water outlet: Hose with 6 mm external diameter / 4 mm internal diameter

Product water

  • Measuring range (Silica SiO₂): 0.3 - 1.2 ppm
  • Limits: 10 limits available in the range of 0.3 to 1.2 ppm

Electrical

  • Power supply: Depending on product configuration: 24V / 115V / 230V, 50 – 60 Hz
  • Instrument protection:
    • 230 – 240 V: T0.1 A
    • 115 V: T0.2 A
    • 24 V: T0.8 A
  • Mains protection for consumers: max. 4 A (N, L)
  • Power consumption: max. 16 VA (without external load)
  • Protection class: I
  • Degree of protection: IP 54
  • Conformity: EN 61000-6-2, EN 61000-6-4, EN 61010-1
  • Current interface: Output of defined values (5, 8, 11, 14, 17, 20 mA) for displaying status and error messages, max. load 500 Ohms
  • Contact load relay: 230V / 4A AC ohm resistive load

Mechanical / hydraulic

  • Dimensions:
    • 364 mm (W) x 314 mm (H) x 138 mm (T)
    • With side pocket: 442 mm (W) x 314 mm (H) x 138 mm (T)
  • Weight: 4350 g