العودة إلى Testomat: hardness controller
Testomat logo

ملخص ورقة البيانات

Testomat 2000 THCL

Engineering summary from PDF text extraction for Testomat 2000 THCL. Verify every value with the OEM datasheet.

Summary

Product identification

The Testomat 2000 THCL® is a process photometer that combines chlorine and hardness measurement in a single device.

Application

  • medical technology (dialysis)
  • corrosion protection
  • protection for reverse osmosis membranes
  • monitoring of softener and chlorination systems for drinking water or swimming pools

Operating limits

Feed water

  • Operating pressure: 1 to 8 bar (15 to 116 psi) or 0.3 to 1 bar (5 to 15 psi) after removing the valve body from the regulator/filter housing
  • Water temperature: 10 to 40 °C (50 to 104°F)

Product water

Measuring range

Parametermg/l (resolution)°dH (resolution)°f (resolution)ppm CaCO₃ (resolution)mmol/l (resolution)
Total Chlorine0.0 - 2.5 (0.1)
Water Hardness0.25 - 2.50 (0.05)0.45 - 4.48 (0.10)4.48 - 44.8 (0.9)0.04 - 0.45 (0.01)

Electrical

  • Power supply: 230–240 VAC, 115 VAC, 24 VAC (all 50–60 Hz)
  • Power consumption: max. 30 VA
  • Protection class: I

Mechanical / hydraulic

  • Degree of protection: IP 65
  • Dimensions (W x H x D): 380 x 480 x 280 mm
  • Weight: approx. 9.5 kg
  • Water inlet: Opaque hose with 6 mm external diameter / 4 mm internal diameter
  • Water outlet: Hose with 6 mm external diameter / 4 mm internal diameter
  • Ambient temperature: 10–45 °C (50 - 104°F)
  • Menu languages: German, English, French, Italian (others upon request)
  • Option: booster pump (see accessories)

Disclaimer: This extracted summary may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.

Official datasheet (PDF)

Lenntech datasheet mirror

أُعد المحتوى من مراجع تقنية عامة للبحث والمقارنة الأولية. لا يغني الملخص عن ورقة بيانات حالية معتمدة من الشركة المصنعة. تحقق من الإصدارات وحدود التصميم قبل الاستخدام.