العودة إلى The AQL Ozone family
Aeroqual logo

ملخص ورقة البيانات

UV-H

Engineering summary from PDF text extraction for UV-H. Verify every value with the OEM datasheet.

Summary

Product identification

The Aeroqual UV-H Photometer Instrument is a portable, single-beam UV photometer designed for ozone measurement.

Operating limits

  • Measurement range: 0-200 ppm (or 0-400 mg/m³) Ozone
  • Environmental Temperature: 0-40°C
  • Relative Humidity: 0-95% RH (non-condensating)

Feed water

  • Sampling method: Brushless DC diaphragm pump
  • Sample flow rate: 0.4 ±0.05 LPM
  • Inlet filter: 5 µm pore size, 37 mm PTFE filter
  • Pneumatic connections: Kynar ¼” compression fittings
  • Wetted materials: Kynar, Teflon PFA, quartz, aluminium, Viton

Product water

  • Precision: <±5% of reading above 0.2 ppm; ±0.01 ppm below 0.2 ppm
  • Resolution: 0.01 ppm
  • T₉₀: < 60s

Electrical

  • Power supply: 24 VDC Switch mode power supply (Input: 90-260V AC; 47-63Hz)
  • Digital interface: RS232 as standard
  • Optional outputs: 0-5V DC analog output (12 bit), setpoint-controlled 30V DC 2A relay

Mechanical / hydraulic

  • Display: 4-line x 20-character VFD
  • Data logging: 2 GB SD card
  • Enclosure: Metal instrument case, dimensions 378 x 236 x 132 mm
  • Software: PC data logging software supplied

Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.

Official datasheet (PDF)

Lenntech datasheet mirror

أُعد المحتوى من مراجع تقنية عامة للبحث والمقارنة الأولية. لا يغني الملخص عن ورقة بيانات حالية معتمدة من الشركة المصنعة. تحقق من الإصدارات وحدود التصميم قبل الاستخدام.